Mass Analyzer · 2 / 3
Separating ions by how long they take to fly · from the basic linear tube to reflectron and spiral designs
In a time-of-flight analyzer, all ions are accelerated with the same energy. Since kinetic energy is E = ½mv², lighter ions end up faster and heavier ions slower, so they separate by arrival time as they fly down a field-free tube. TOF does not throw any ions away — it detects them all, just at different times.
On this page: Linear · Reflectron · Spiral
The relationship is t ∝ √(m/z) — a mass four times larger takes twice as long to arrive. But real ions of the same mass leave with a slight spread of energies, so they arrive a little apart and the peak is broadened. The reflectron fixes exactly this.
The mirror does not make different masses arrive together (that is impossible and would break TOF). It brings ions of the same mass but different energy back into step — the extra distance inside the mirror exactly offsets the extra speed. The result is much higher resolution.